TB 9-6625-2274-50
TEKTRONIX, TYPE CG-5011
6. Preliminary
Instructions
a. The
instructions outlined in
paragraphs 6 and
7 are preparatory to the
calibration
process. Personnel should become familiar with the entire bulletin before beginning the
calibration.
b. Items of equipment used in this procedure are referenced within the text by common
c. Unless otherwise specified, verify the results of each test and, whenever the test
requirement is not met, take corrective action before continuing with the calibration.
Adjustments required to calibrate the TI are included in this procedure. Additional
maintenance information is contained in the manufacturer's manual for this TI.
d.
When indications specified in
paragraphs 8 through 13 are not within
tolerance,
perform Section IV, Adjustment Process for Programmable Calibration Generator,
Do not perform if all other parameters are within tolerance.
e. Unless otherwise specified, all controls and control settings refer to the TI.
f. Unless otherwise specified, all connections referring to OUTPUT are made through
the output cable supplied with the TI.
WARNING
HIGH VOLTAGE is used or exposed during the performance of
personnel fail to observe safety precautions. REDUCE
OUTPUT(S) to minimum after each step within the
performance check where applicable.
a. Connect TI to a 115 V ac source.
b. Energize TI and allow at least 20 minutes for warm-up and stabilization.
a. Performance Check
(1) Connect TI OUTPUT to multimeter voltage input. Set multimeter to measure
autorange DCV.
(2) Press pushbuttons as listed in (a) through (e) below:
(a) AMPLITUDE MODE AMPL MODE VOLT.
(b) AMPLITUDE MODE FREQ
to highlight +DC.