TB 9-6625-2274-50
(c) OUTPUT USE FOR 5O Ω LOAD off.
(d) VARIABLE ON off.
(e) OUTPUT ON on.
(3) Adjust UNITS/DIV dial to 20 V/D on display and press AMPLITUDE MODE
NUMBER OF DIV (MULTIPLIER)
pushbutton to display X l0 DIV. Multimeter will
indicate between +199.50 and +200.50 V dc.
(4) Press AMPLITUDE MODE FREQ
pushbutton to highlight -DC.
Multimeter will indicate between -199.50 and -200.50 V dc.
(5) Press AMPLITUDE MODE FREQ
pushbutton to highlight +DC and
repeat technique of (3) and (4) above for TI settings listed in table 3.
Test instrument
Multimeter indications
(V dc)
NUMBER OF DIV
(MULTIPLIER)
UNITS/DIV
pushbutton
dial display
display settings
settings
(DIV)
(V/D)
Max
Min
10
10
99.75
V
100.25
V
10
2
19.95
V
20.05
V
10
1
9.975
V
10.025
V
1
8
7.98
V
8.02
V
1
6
5.985
V
6.015
V
1
5
4.9875 V
5.0125
V
1
4
3.99
V
4.01
V
1
3
2.9925 V
3.0075
V
.5
5
2.4938 V
2.5062
V
.5
3
1.4963 V
1.5037
V
.1
1
99.75
mV
100.25
mV
(6) Position controls as listed in (a) through (d) below:
(a) Press AMPLITUDE MODE FREQ
pushbutton to highlight -DC.
(b) Press AMPLITUDE MODE NUMBER OF DIV
pushbutton to display
X1 DIV.
(c) Press VARIABLE ON pushbutton on.
(d) Adjust UNITS/DIV dial to display 10 V/D.
(7) Adjust UNITS/DIV VAR dial to 5.0% LOW on display. Multimeter will indicate
between -10.5000 and -10.5526 V dc.
(8) Press AMPLITUDE MODE FREQ
pushbutton to highlight +DC.
Multimeter will indicate between + 10.5000 and +10.5526 V dc.
(9) Press AMPLITUDE MODE FREQ
pushbutton to highlight -DC and repeat
technique of (7) and (8) above for TI settings listed in table 4.
(10) Press OUTPUT ON pushbutton off.
6