TB 9-6625-2274-50
b. Apply power and allow at least 20 minutes for warm-up and stabilization.
a. Press TIMING MODE MARKERS pushbutton and adjust UNITS/DIV dial to
display .5 s.
b. Set multimeter for dc volts and autorange. Connect positive lead of multimeter to TI
c.
Adjust A5Rl004 (fig. 1) for a
multimeter indication of
+5.15 V dc
0.05 V
(R).
d.
Remove multimeter connections and
connect 20k, 2W, 5%
resistor to
OUTPUT.
e.
Connect multimeter positive lead to
A5TP1702 (fig. 1) and
negative lead to
f. Position controls as listed in (1) through (5)below:
(1) Press AMPLITUDE MODE AMPL MODE VOLT pushbutton.
(2) Press AMPLITUDE MODE FREQ
pushbutton to highlight +DC.
(3) Press AMPLITUDE MODE NUMBER OF DIV (MULTIPLIER)
pushbutton to display X 10 DIV.
(4) Adjust UNITS/DIV dial to display 20 V/D.
(5) Press OUTPUT ON pushbutton on.
g.
Adjust A5R1106 (fig. 1) for a
multimeter indication of
+237 V dc 2 V
(R).
h. Press OUTPUT ON pushbutton off. Disconnect resistor from OUTPUT.
i. Adjust UNITS/DlV dial to display lV/D and press AMPLITUDE MODE NUMBER
OF DIV (MULTIPLIER)
pushbutton to display X 1 DIV.
j.
Connect multimeter positive lead to A5TP1102 (fig. 1).
k.
Adjust A5R1104 (fig. 1) for a
multimeter indication of
+15.00 V dc
02 V
(R).
l.
m. Set cal/normal switch A9S1061 (fig. 1) to the
cal position (down). TI
display will
indicate DAC 0.