d. Perform risetime measurement technique using digitizing oscilloscope. Adjust
A7C1621 (fig. 1) until digitizing oscilloscope indicates a risetime less than 1.3 ns and
minimum aberrations (R).
e. Press OUTPUT ON pushbutton off and remove digitizing oscilloscope connections.
a. Connect TI output to oscilloscope CH1.
b. Position controls as listed in (l) through (7) below:
(1) Press AMPLITUDE MODE AMPL MODE EDGE pushbutton.
(2) Press AMPLITUDE MODE FREQ
pushbutton to highlight 1 kHz.
(3) Press AMPLITUDE MODE NUMBER OF DIV (MULTIPLIER)
pushbutton to display X 8 DIV.
pushbutton.
(4) Press AMPLITUDE MODE EDGE POLARITY
(5) Press VARIABLE ON pushbutton off.
(6) Adjust UNITS/DIV dial to display 2 V/D.
(7) Press OUTPUT ON pushbutton on.
c. Acquire signal and
set oscilloscope amplitude to 2
V/D. Adjust oscilloscope variable
control for exactly 6 vertical divisions of trace separation.
d. Position controls as listed in (1) through (6) below:
(1) Press AMPLITUDE MODE AMPL MODE VOLT pushbutton.
(2) Press AMPLITUDE MODE NUMBER OF DIV (MULTIPLIER)
pushbutton to display X 8 DIV.
(3) Press AMPLITUDE MODE FREQ
to highlight 1 kHz.
(4) Press VARIABLE ON pushbutton on.
(5) Adjust UNITS/DIV dial to display 2 V/D.
(6) Press OUTPUT ON pushbutton on.
e. Adjust TI VAR dial until oscilloscope indicates exactly 6 divisions of trace
separation. Do not change once set.
f. Adjust UNITS/DlV dial to display .2V/D and press AMPLITUDE MODE
NUMBER OF DIV (MULTIPLIER)
pushbutton to display X 6 DIV.
g.
Set oscilloscope amplitude to
200 mV/D and
adjust variable control until
display
indicates exactly 5 vertical divisions of trace separation.
h. Position controls as listed in (1) through (4) below:
(1) Press AMPLITUDE MODE AMPL MODE EDGE pushbutton.
(2) Press AMPLITUDE MODE NUMBER OF DIV (MULTIPLIER)
pushbutton to display X 6 DIV.
(3) Adjust UNITS/DIV dial to display .2 V/D.