TB 9-6625-2274-50
(4) Press OUTPUT ON pushbutton on.
i.
Adjust A7R1408 (fig. 1) until
oscilloscope displays exactly 5
vertical divisions of
trace
separation.
j. Adjust UNITS/DIV dial to display 2 V/D and press AMPLITUDE MODE FREQ
pushbutton to highlight 100 kHz.
k. Acquire signal using oscilloscope and
set amplitude to 2
V/D.
Perform risetime
technique using oscilloscope.
Adjust A7R1527 (fig. 1) until oscilloscope displays a risetime of ≤40 ns.
l.
m. Press OUTPUT ON pushbutton off and connect X10 probe between
oscilloscope and TI.
n. Position controls as listed in (1) through (6) below:
(1) Press AMPLITUDE MODE AMPL MODE EDGE pushbutton.
(2) Press AMPLITUDE MODE NUMBER OF DIV (MULTIPLIER)
pushbutton to display X 5 DIV.
(3) Press AMPLITUDE MODE FREQ
pushbutton to highlight 1 kHz.
(4) Adjust UNITS/DIV dial to display 20 V/D.
(5) Press VARIABLE ON pushbutton off.
(6) Press OUTPUT ON pushbutton on.
o. Acquire signal using oscilloscope and
set amplitude to 2
V/D. Adjust oscilloscope
variable control until display indicates exactly 4 vertical divisions of trace separation.
p. Position controls as listed in (1) through (6) below:
(1) Press AMPLITUDE MODE AMPL MODE VOLT pushbutton.
(2) Press AMPLITUDE MODE NUMBER OF DIV (MULTIPLIER)
pushbutton to display X 5 DIV.
(3) Press AMPLITUDE MODE FREQ
pushbutton to highlight 1 kHz.
(4) Adjust UNITS/DIV dial to display 20 V/D.
(5) Press VARIABLE ON pushbutton on.
(6) Press OUTPUT ON pushbutton on.
q. Adjust TI VAR dial for exactly 4 vertical divisions of trace separation on
oscilloscope. Do not change once set.
r. Adjust UNITS/DIV dial to display 5 V/D and press AMPLITUDE MODE
NUMBER OF DIV (MULTIPLIER)
pushbutton to display X 4 DIV.
s.
Adjust oscilloscope amplitude to
200 mV/D and
adjust variable control for 6
vertical
divisions of trace separation.
t. Position controls as listed in (1) through (4) below:
(1) Press AMPLITUDE MODE AMPL MODE EDGE pushbutton.
21