TB 9-6625-2274-50
k. Counter will indicate a display of 5000 kHz .25 kHz. If TI does not indicate within
limits specified, continue with paragraph m below. If TI indicates within limits specified,
continue with paragraph n below.
l. Remove screw covering adjustment hole on the
top of TCXO
OPT01 (fig. 1) case.
Note the marking on the TCXO OPT01 (fig. 1) case. Adjust TCXO OPT01 (fig. 1) frequency
trimmer for a frequency counter display equal to frequency marked on TCXO OPT01 (fig. 1)
case (R).
m. Replace screw covering adjustment hole.
n. Press TRIGGER OUTPUT ON pushbutton off and disconnect equipment setup.
PULSE HEAD TEKTRONIX, TYPE 015-0611-00
23. Preliminary
Instructions
a. The
instructions outlined in
paragraphs 23 and
24 are preparatory to the
calibration
process. Personnel should become familiar with the entire bulletin before beginning the
calibration.
b. Items of equipment used in this procedure are referenced within the text by common
c. Unless otherwise specified, verify the result of each test and, whenever the test
requirement is not met, take corrective action before continuing with the calibration.
Adjustments required to calibrate the TI are included in this procedure. Additional
maintenance information is contained in the manufacturer's manual.
d. Unless otherwise specified, all controls and control settings refer to the TI.
WARNING
HIGH VOLTAGE is used or exposed during the performance of
this calibration. DEATH ON CONTACT may result if
personnel fail to observe safety precautions. REDUCE
OUTPUT(S) to minimum after each step within the
performance check where applicable.
a. Remove TI bottom protective cover, as required, for adjustments.
b. Connect TI to programmable calibration generator OUTPUT.
c.
Connect programmable calibration generator to a
115 V ac
source.
warm-up and stabilization.