TB 9-6625-2274-50
a. Performance Check
(1) Acquire signal using digitizing oscilloscope and set sweep speed to 1 ns/D.
Adjust delay to center waveform on vertical center graticule.
(2) Set digitizing oscilloscope sensitivity to 10 mV/D and adjust offset to place the
top of waveform on center horizontal graticule.
(3) Adjust digitizing oscilloscope offset and delay until waveform is displayed as
(4) Perform Delta V measurement technique, using oscilloscope, by placing one
marker on the upper peak of the aberration and one marker on center graticule. Digitizing
oscilloscope Delta V (peak aberrations will indicate <3 percent of amplitude recorded
in 25 a (5) above.
(5) Adjust one Delta V marker of oscilloscope to the lower peak of the aberration
and one to center horizontal graticule.
(6) Digitizing oscilloscope Delta V (peak aberration) will indicate <3 percent of
amplitude recorded in 25 a (5) above.
(7) Adjust digitizing oscilloscope Delta V markers to the upper and lower
aberrations.
(8) Digitizing oscilloscope Delta V (peak-to-peak aberration) will indicate <4 percent
of amplitude recorded in 25 a (5) above.
(9) Maintain setup and continue with next performance check.
b. Adjustments
(1) Press OUTPUT ON pushbutton off and remove X10 attenuator. Connect
OUTPUT to digitizing oscilloscope INPUT/TDR OUTPUT using PULSE HEAD and X5
attenuator.
(2) Press pushbuttons as listed in (a) through (g) below:
(a) AMPLITUDE MODE AMPL MODE FAST EDGE.
(b) AMPLITUDE MODE EDGE POLARITY .
(c) AMPLITUDE MODE FREQ
to highlight l00 kHz.
25