Figure 1. A3S1 service switch.
Instrument Accuracy Test

Calibration Procedure For Power Meter, Model 438A And 438Aopt002 Page Navigation
  1    2    3    4    5  6  7    8    9    10    11  
img
. .
TB 9-6625-2278-24
Figure 2. A5 component location.
(12) Disconnect multimeter from TI.
(13) Press LINE switch to OFF position.
(14) Place all four switches of service switch A3S1 (fig. 1) into NORMAL position.
7