Instrument Accuracy Test - Continued
Power Reference Level

Calibration Procedure For Power Meter, Model 438A And 438Aopt002 Page Navigation
  5    6    7    8    9  10  11    12    13    14    15  
img
. . .
TB 9-6625-2278-24
(18) Set range calibrator RANGE switch to 10 mW and adjust A4R27 (fig. 3) for a TI
display indication between 9.99 and 10.01 mW (R)
(19) Set range calibrator RANGE switch to 100 mW and adjust A4R35 (fig. 3) for a
TI display indication between 99.9 and 100.1 mW (R).
Figure 3. A4 board - component location.
11