SECTION III. CALIBRATION PROCESS
Capacitance Measuring Section Accuracy

Calibration Procedure For Fuel Quantity System Test Set Air Model Psd 60-1Af Page Navigation
  1    2  3  4    5    6    7    8  
img
. . .
TB 9-6625-2285-24
Figure 1. Front panel.
8. Capacitance Measuring Section Accuracy
a. Performance Check
(1) Set SELECT switch to TU.
(2) Set FUNCT switch to MEASURE EXT.
(3) Set variable capacitor to 35 pf.
NOTE
Over range is indicated when display indicates :0:0:00.
(4) Connect variable capacitor 115 output terminal to TI TANK UNITS HIZ
terminal using coaxial lead.