TB 9-6625-2285-24
(10) Connect the center pins of TI INDICATOR LOZ to TANK UNITS LOZ with a
short banana to banana jumper lead and 2 pin to banana adapters.
(11) Set SELECT switch to TU.
(12) Set FUNCT switch to SIM TU ONLY.
(13) Measured capacitance as indicated on TI digital readout will be between 19.8
and 20.2 pf.
(14) Set SELECT switch to COMP.
(15) Move connection from TI TANK UNITS LOZ to chassis.
(16) Set FUNCT switch to SIM TU & COMP.
(17) Repeat (13) above.
(18) Set FUNCT switch to MEASURE INT and remove all cables from TI.
(19) Press and release + pushbutton below 10 digits of COMP decade to increment
decade to 01 and verify TI display increases by approximately 10.
(20) Repeat technique of (19) above for 10 digits 2 through 9.
(21) Repeat technique of (19) above for 100 digits 1 through 9 and verify TI display
increases by approximately 100.
(22) Set SELECT switch to TU.
(23) Press and release + pushbutton below 10 digits on TU decade and verify TI
display increases by approximately 10.
(24) Repeat technique of (23) above for digits 2 through 9.
(25) Repeat technique of (23) above for 100 digits and verify TI display increases by
approximately 100.
NOTE
Press PF EXTENDED RANGE pushbutton for TI display
indications above 1,999.
(26) Repeat technique of (23) above for 1000 digits 1 through 7 and verify TI display
increases by approximately 1000.
b. Adjustments. No adjustments can be made.
a. Performance Check
(1) Set SELECT switch to LOZ-HIZ.
(2) Set FUNCT switch to MEASURE EXT.