SECTION I. IDENTIFICATION AND DESCRIPTION
SECTION III. CALIBRATION PROCESS

Calibration Procedure For Oscilloscope Types 2465A And 2465 Page Navigation
  1    2  3  4    5    6    7    8  
img
. . . . . .
TB 9-6625-2294-24
Table 1. Calibration Description - Continued
Test instrument parameters
Performance specification
Trigger level readout
Accuracy:  Within [3% of setting + 3% of p-p signal
+ 0.2 division + 0.5 mV +(0.5 mV x probe
attenuation factor)]
Trigger sensitivity CH 1 and CH 2
Dc coupled
0.35
division
Dc
to
50
MHz
1.0
division
at
300
MHz
1.5
division
at
500
MHz
Ac coupled
0.35
division
60 Hz
to
50
MHz
1.0
division
at
30
MHz
1.5
division
at
500
MHz
Horizontal timing
Range: 500 ms/div to 5 ns/div extended to 0.5 ns with X10 mag
A sweep
Accuracy:  500 and 200 ms (1.2% of time interval + 0.6% of FS)
100 ms to 5 ns (0.7% of time interval + 0.6% of FS)
X10 mag: (1.2% of time interval + 0.6% of FS)
B sweep
Range:  50 ms/div to 5 ns/div extended to 0.5 ns with X10 mag
1.0% of time interval + 0.6% of FS
Accuracy:
X10 mag: 1.5% of time interval + 0.6% of FS
(0.5% of time interval + 0.3% of FS)
Accuracy:
X10 mag: (1% of time interval + 0.3% of FS)
Delta time with cursors
(0.3% of time interval + 0.1% of FS + 200 ps)
Accuracy:
(0.8% of time interval + 0.1% of FS + 200 ps)
Delta time with sweep delay
X10 mag:
(0.8% of time interval + 0.3% of FS)
Accuracy:
Delta time with cursors on
(1.3% of time interval + 0.3% of FS)
X10 mag:
B sweep
(0.3% of delay setting + 0.6% of FS) +0 to -25 ns
Accuracy:
Delay sweep
Bandwidth1
DC to 350 MHz
0.4V 1% into 1 M
Calibrator output voltage
load
1Not tested below 50 kHz.
SECTION II
EQUIPMENT REQUIREMENTS
4. Equipment Required. Table 2 identifies the specific equipment to be used in this
calibration procedure.  This equipment is issued with Secondary Transfer Calibration
Standards Sets AN/GSM-286, AN/GSM-287 or AN/GSM-705. Alternate items may be used
by the calibrating activity. The items selected must be verified to perform satisfactorily
prior to use and must bear evidence of current calibration. The equipment must meet or
exceed the minimum use specifications listed in table 2. The accuracies listed in table 2
provide a four-to-one ratio between the standard and TI.  Where the four-to-one ratio
cannot be met, the actual accuracy of the equipment selected is shown in parenthesis.
3