• Home
  • Download PDF
  • Order CD-ROM
  • Order in Print
Table 1. Calibration Description
Equipment Setup

Calibration Procedure For Oscilloscope Types 2465A And 2465 Page Navigation
  1    2    3  4  5    6    7    8    9  
img
. . . . . . . . . . . .
TB 9-6625-2294-24
5. Accessories Required.  The accessories required for this calibration are common
usage accessories issued as indicated in paragraph 4 above and are not listed in this
calibration procedure.  The following peculiar is also required for this calibration:
Standardizer, 5 80 pF; BNC plug to BNC jack (7916146).
Table 2. Minimum Specifications of Equipment Required
Manufacturer and model
Common name
Minimum use specifications
(part number)
MULTIMETER
Range: 0 to 20 V
Fluke, Model 8840A/AF05
Accuracy:  0.025%
(AN/GSM-64D)
OSCILLOSCOPE
Used in CAL 01 adjustment procedure
Agilent, OS-303/G (OS-303/G)
OSCILLOSCOPE
Volts out:
Fluke, Model 5820A-5C-GHZ
CALIBRATOR
Range: 10 mV to 20 V
(5820A-5C-GHZ)
Accuracy: 0.5%
Time markers:
Range: 0.5 ns to .5 s
Accuracy: 0.175%
Pulses:
Risetime: 0.225 ns
Sine wave frequency:
Range: 1 Hz to 500 MHz
SECTION III
CALIBRATION PROCESS
6. Preliminary Instructions
a. The instructions outlined in paragraphs 6 and 7 are preparatory to the calibration
process. Personnel should become familiar with the entire bulletin before beginning the
calibration.
b. Items of equipment used in this procedure are referenced within the text by common
name as listed in table 2.
c. Unless otherwise specified, verify the results of each test and, whenever the test
requirement is not met, take corrective action before continuing with the calibration.
Adjustments required to calibrate the TI are included in this procedure.  Additional
maintenance information is contained in the manufacturer's manual for this TI.
d. When indications specified in paragraphs 7 through 12 are not within tolerance,
perform SECTION IV, MENU CALIBRATION PROCESS. After adjustments are made,
repeat paragraphs 7 through 12. Do not perform SECTION IV if all other parameters are
within tolerance.
e. Unless otherwise specified, all controls and control settings refer to the TI.
4


Privacy Statement
Press Release
Contact

© Copyright Integrated Publishing, Inc.. All Rights Reserved.