SECTION I. IDENTIFICATION AND DESCRIPTION

Calibration Procedure For Oscilloscope Types 2465A And 2465 Page Navigation
1  2    3    4    5    6  
img
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
*TB 9-6625-2294-24
DEPARTMENT OF THE ARMY TECHNICAL BULLETIN
CALIBRATION PROCEDURE FOR
OSCILLOSCOPE
TEKTRONIX, TYPES 2465A AND 2467
Headquarters, Department of the Army, Washington, DC
26 November 2007
Distribution Statement A: Approved for public release; distribution is unlimited.
REPORTING OF ERRORS AND RECOMMENDING IMPROVEMENTS
You can improve this manual. If you find any mistakes or if you know of a way to improve these
procedures, please let us know.  Mail your letter or DA Form 2028 (Recommended Changes to
Publications and Blank Forms) directly to: Commander, U.S. Army Aviation and Missile Command,
ATTN: AMSAM-MMC-MA-NP, Redstone Arsenal, AL 35898-5000. A reply will be furnished to
you.
You  may also send in your comments electronically to our E-mail address:
2028@redstone.army.mil or by fax 256-842-6546/DSN 788-6546.  For the World Wide Web use:
https://amcom2028.redstone.army.mil. Instructions for sending an electronic 2028 can be found at
the back of this manual.
Paragraph
Page
SECTION
I.
IDENTIFICATION AND DESCRIPTION
Test instrument identification ...........................
1
2
Forms, records, and reports ...............................
2
2
Calibration description.......................................
3
2
II.
EQUIPMENT REQUIREMENTS
Equipment required ...........................................
4
3
Accessories required ...........................................
5
4
III.
CALIBRATION PROCESS
Preliminary instructions ....................................
6
4
Equipment setup.................................................
7
5
Vertical gain ........................................................
8
6
Triggering .........................................................
9
12
Timing
.........................................................
10
15
Bandwidth .........................................................
11
22
Calibrator  .........................................................
12
24
Final procedure ...................................................
13
25
_________
*This bulletin supersedes TB 9-6625-2294-35, dated 10 March 2003, including all changes.