(2) Connect multimeter HI to TI J119 (fig. 1) pin 4 and LO to chassis ground.
If
multimeter does not indicate between +9.99 and +10.01 V dc, perform b below.
b. Adjustments. Adjust VOLT REF ADJ R1292 (fig. 2) for a multimeter indication of
10.00 V dc (R).
Figure 1. A1 main board - adjustment locations.
CRT
TERMINATION
R1501
R2010
VOLT REF ADJ
R1292
Figure 2. Adjustment locations.