TB 9-6625-2294-35
(a) Set CH 1 VOLTS/DIV switch as listed in first row of table 11.
(b) Set oscilloscope calibrator VOLTAGE output as listed in first row of table
(7) Perform steps (a) through (d) below:
(a)
Pull SEC/DIV knob out.
(b)
Press A/B TRIG pushbutton for B TRIGGER indication
(c)
Press TRIGGER MODE pushbutton for B TRIG AFT DLY.
(d)
Adjust TRIGGER LEVEL control for a stable display.
(8) Adjust ∆REF OR DLY POS control for a delay readout of 0.000 ms.
(9) Alternately adjust TRIGGER LEVEL control for most positive voltage or most
negative voltage that produces an intensified point on waveform display when
selecting + or - SLOPE. If B TRIGGER level readout indication is not within limits
given in + peak column or - peak column of table 11, perform b below.
(10) Repeat technique of steps (6) through (9) above for each remaining row listed in
Table 11. Trigger Level 1 MΩ
TRIG AFT DLY Input Coupling
Test instrument trigger level
Test instrument
calibrator
readout indications
-peak
VOLTS/DIV
VOLTAGE
+ peak
switch settings
output settings
Min
Max
Min
Max
5
mV
20
mV
16.8
mV
23.2
mV
-2.6
mV
2.6
mV
10
mV
50
mV
44
mV
56
mV
-4.5
mV
4.5
mV
20
mV
0.1
V
89
mV
111
mV
-8.0
mV
8.0
mV
50
mV
0.2
V
178
mV
222
mV
-16
mV
16
mV
100
mV
0.5
V
450
mV
550
mV
-35
mV
35
mV
200
mV
1.0
V
0.90
V
1.10
V
-70
mV
70
mV
500
mV
2.0
V
1.78
V
2.22
V
-.16
V
.16
V
1.0
V
5.0
V
4.50
V
5.50
V
-.35
V
.35
V
2.0
V
10.0
V
9.0
V
11.0
V
-0.7
V
0.7
V
5.0
V
20.0
V
17.8
V
22.2
V
-1.6
V
1.6
V
(11) Press corresponding pushbuttons for indications listed in (a) through (f) below:
(a)
20 MHz BW LIMIT off.
(b)
TRIGGER SLOPE to + (plus).
(c)
TRIGGER COUPLING to AC.
(d)
A/B TRIG (off).
(e)
TRIGGER MODE to AUTO LVL.
(f)
TRIGGER SOURCE to VERT CH 1.
(12) Set CH 1 VOLTS/DIV switch to 100 mV and set A SWP SEC/DIV switch to 10
ms (knobs locked; pushed in).
14