TB 9-6625-2296-24
(15)Stop Bits .........................................
1
(16)Parity ..............................................
None
(17)Handshake .....................................
Xon/Xoff
(18)Echo ................................................
Off
ff. Press ESC (F6) twice then TERM (F5).
gg. Connect Serial Interface (null modem) Cable Assembly between Test Set RS-232
connector and Test Adapter RS-232 Connector. Connect Test Fixture TF-119 to Test
Adapter.
hh. Press SHIFT, T, E, S, T, C, O, N, N, SHIFT and ENTER Keys.
ii. Press 1 on DATA ENTRY Keypad.
NOTE
Test is performed automatically with user instructions given as needed.
jj. Press key sequence listed in (1) through (4) below:
(1) MODE RCVR ..............................
Receiver screen
(2) More (F6) .....................................
Additional Selections
(3) DATA ENTRY 1 2 1 . 1 M ..........
121.1 MHz
(4) Verify RF IN indicates ANT; if not, press Ant (F3) key.
ll.
Set signal generator for a 121.1 MHz, -30 dBm output.
mm. Press key sequence listed in (1) through (6) below:
(1) MODE MTRS ..............................
Meter Menu
(2) AUX (F6) ......................................
Auxiliary Functions Menu
(3) DATA ENTRY 1..........................
(4) DATA ENTRY 7..........................
Signal Meter
(5) DATA ENTRY ENTER..............
Enter to Cal
(6) DATA ENTRY 1 0 0 ENTER ....
Correction Factor
nn. Press Ret (F5) (Return), MODE RCVR, DATA ENTRY SHIFT, DATA ENTRY
Y (Backup Cal Data ? y/n) keys.
oo. Disconnect signal generator from TI.
14