TB 9-6625-2296-24
NOTE
Before starting calibration, make sure TI has the latest
software. In the upper right portion of the screen the software
version will be listed (e.g. TS-4317 V. X.XX, PL-1536/GRM V.
X.XX, etc.). The software may be obtained at the USATA
website Publications > Masterlist > TB 9 > TB 9-6625-2296-24.
Update software if applicable (reprogramming does not affect
calibration).
NOTE
If no auxiliary box is connected to TI rear panel SCSI, TI SCSI
must be disabled in hh below prior to performing self test in ii
below. Do not perform hh below if auxiliary box is connected to
TI rear panel SCSI.
s. Press key sequence listed in (1) through (5) below:
(1) MODE MTRS ..............................
Meters Menu
(2) AUX (F6) ......................................
Auxiliary Functions Menu
(3) DATA ENTRY 5..........................
Configure submenu
(4) DATA ENTRY 3..........................
Configure SCSI
(5) DATA ENTRY ENTER..............
Operation Mode off
t. Press key sequence listed in (1) through (14) below:
(1) MODE MTRS ..............................
Meters Menu
(2) AUX (F6) ......................................
Auxiliary Functions Menu
(3) DATA ENTRY 1..........................
Calibration Menu
(4) Cal msg (F1) ................................
Cal message Menu
(5) Enable Messaging .....................
Toggle to off
(6) ESC (F6).......................................
Cal message Menu
(7) AUX (F6) ......................................
Auxiliary Functions Menu
(8) MEMORY RCL ...........................
Recall Parameters Menu
(9) FIELD SELECT ......................
Highlight 10. Factory Defaults
(10)DATA ENTRY ENTER..............
Factory Defaults recall
NOTE
Assure highlighted area indicates YES before proceeding.
10