TB 9-6625-2296-24
(11) DATA ENTRY ENTER ...........
Factory Defaults restored
(12) AUX (F5) ....................................
Auxiliary Functions Menu
(13) DATA ENTRY 4 .......................
Self Test
(14) DATA ENTRY ENTER ...........
Test Running
NOTE
When self test is completed, ensure P is displayed next to
Self Test; otherwise, refer to TM for maintenance action.
NOTE
The following test (steps u and v below) is performed for the
Perform w through ii below for the
AN/GRM-122.
AN/GRM114B. For calibration of the TS-4317/GRM proceed to
jj below.
NOTE
Using cables provided by the customer connect the TS-
4317/GRM DUPLEX, EXT MOD, SINAD BER, AUDIO,
DEMOD, DMM and SCOPE connectors to the corresponding
PL-1536/GRM connectors.
u. Press key sequence listed in (1) through (15) below:
(1) MODE MTRS .................................
Meters Menu
(2) AUX (F6) .........................................
Auxiliary
(3) MEMORY RCL ..............................
Recall
(4) FIELD SELECT .........................
Highlight 10. Factory Defaults
(5) DATA ENTRY ENTER .................
Factory Defaults recall
NOTE
Assure highlighted area indicates YES before proceeding.
(6) DATA ENTRY ENTER .................
Factory Defaults restored
(7) MODE RCVR .................................
Receiver Menu
(8) Sp Tst (F5)......................................
SINCGARS Verification Test Menu
(9) More "x" (F6) .................................
Until BIT 1 (F2) and BIT 2 (F3) are
displayed
(10)BIT 1 (F2) .......................................
BIT Level 1 Test
(11)AUTO...............................................
OPER Mode (Auto Mode)
(12)GO ....................................................
BIT Level 1 Test begins
11