TB 9-6625-2296-35
Table 2. Minimum Specifications of Equipment Required - Continued
Manufacturer and model
Common name
Minimum use specifications
(part number)
MEASURING RECEIVER
Power measurement:
Hewlett-Packard, Model 8902A
Flatness: 500 kHz to 999 MHz
(8902A) w/sensors,
0 to +30 dB 0.103 dB
Hewlett-Packard, Model 11722A
Power:
(11722A)
0 to -110 0.375 dB
-110 to -127 0.625 dB
Accuracy: 0.5 dB
FM: Range: 10 to 90 kHz
Accuracy: 1% <15 kHz mod rate
1.75 >15 kHz mod rate
AM: Range: 0.1% 90%
Accuracy: 1.25
Range: 50 Hz
(OS-303/G)
Accuracy: <125 ns risetime
Fluke Model 5820A-5C-GHz
Voltage:
(5820A-5C-GHz)
Range: 4 mV to 100 V dc
CALIBRATOR
Accuracy: 2.0%
Leveled sine wave:
Markers:
Range: 2 s to 100 ms/div
Accuracy: 0.25%
Weinschel, Model 1870A
POWER SPLITTER
Insertion loss -6dB
(7916839)
Output tracking
between ports: 10 MHz to 1 GHz:
0.15 dB
Biddle-Gray, Model 71-631
Accuracy: 1%
(7910328)
Bird, Model 4421 (4421) w/
RF POWER METER
Directional power sensors,
Output: 7.5 to 85 W
Accuracy: ( 3 %)
Model 4021 and 4022 (4021and
4022)
SPECTRUM ANALYZER
Frequency: 1 to 999.999 MHz
(AN/USM-677)
Noise Floor: <50 dB
(SG-1207/U)
Frequency: 1 to 900.15 MHz
Resolution: 1 Hz
Accuracy: 1.125 ppm
Output: 0 to -40 dB
2 to 5 V
Mod Freq: 100 Hz to 19 kHz
FM: 1 to 90 kHz
AM: 30 to 90 %
ITEM DELETED
CHANGE 3