TB 9-6625-2296-35
Recall Parameters Menu
(c) MEMORY RCL .....................
(d) FIELD SELECT ⇑ ................
Highlight 10. Factory Defaults
Factory Defaults recall
(e) DATA ENTRY ENTER ........
NOTE
Assure highlighted area indicates YES before proceeding.
Factory Defaults restored
(f) DATA ENTRY ENTER ........
set to 150 Ω.
(3) Press key sequence listed in (a) through (k) below:
AF GEN Screen
(a) MODE AF GEN.....................
AF Gen Menu
(b) SETUP ...................................
AF Output Setup
(c) DATA ENTRY 5 ....................
Func Gen Out Level
(d) DATA ENTRY 4 ....................
2 V setting
(e) DATA ENTRY 2 ENTER .....
AF Gen Menu
(f) ESC (F6) ................................
AF Gen # 1 Setup
(g) DATA ENTRY 1 ....................
Func Gen 1 On
(h) DATA ENTRY 1 ....................
Freq
(i) DATA ENTRY 2 ....................
40 kHz setting
(j)
DATA ENTRY 4 0 k .............
Return
(k) Ret (F5) ..................................
(4) Frequency counter will indicate within limits specified in table 4 for TI AF Gen
frequency 40 kHz setting.
(5) Press FIELD SELECT ⇑ or ⇓ key to select GEN 1 AF.
(6) Press DATA ENTRY ENTER key and repeat technique of (3) (j) above for each
remaining TI AF Gen frequency setting listed in table 4. Frequency counter indications will
Table 4. AF
Generator Frequency/Distortion
Audio analyzer
Test instrument
distortion
AF Gen frequency
indications
settings (Hz)
Min
Max
40
k
39.9600 kHz
40.040
kHz
<1.0%
20
k
19.980
kHz
20.020
kHz
<1.0%
10
k
9.990
kHz
10.010
kHz
<1.0%
1.1
k
1.0989 kHz
1.1011
kHz
<1.0%
1.0
k
999.0000 Hz
1.0010
kHz
<0.7%
7001
699.3000 Hz
700.7000
Hz
<0.7%
101
9.9900 Hz
10.0100
Hz
<1.0%
1Press
DATA ENTRY ENTER key for Hz.
(7) Press key sequence listed in (a) through (g) below:
11