TB 9-6625-2296-35
(8) Press DATA ENTRY ENTER key to set GEN 1 Off. Press FIELD SELECT ⇓
key to select GEN 2 and press DATA ENTRY ENTER key to set GEN 2 On.
(9) If multimeter does not indicate 0.000 .01 V dc adjust A24R22 (fig. 2) as close as
possible.
(10) Press DATA ENTRY ENTER key to set GEN 2 Off. Press FIELD SELECT ⇑
key to select GEN 1 and press DATA ENTRY ENTER key to set GEN 1 On.
(11) Press FIELD SELECT ⇓ key to select GEN 1 WAVE, adjust DATA SCROLL
knob to select Sine, and press DATA ENTRY ENTER key. Set multimeter to measure
volts ac and record multimeter indication (approximately 3.1 V rms).
(12) Press FIELD SELECT ⇑ key to select GEN 1 and press DATA ENTRY
ENTER key to set GEN 1 Off.
(13) Press FIELD SELECT ⇓ key to select GEN 2 WAVE, adjust DATA SCROLL
knob to select Sine, and press DATA ENTRY ENTER key. Press FIELD SELECT ⇑ key
to select GEN 2 and press DATA ENTRY ENTER key to set GEN 2 On.
(14) Adjust A24R85 (fig. 2) for a multimeter indication recorded in (11) above 5 mV (R).
(16) Press MODE MTRS key, AUX (F6) key, DATA ENTRY key 1 (Calibrations),
and press FIELD SELECT ⇓ key to select 15. Func Gen Level. Press DATA ENTRY
ENTER and DATA ENTRY 1 keys (199.9 mV calibration).
(17) Use DATA SCROLL controls to adjust multimeter indication for 199.0 1 mV
ac and press DATA ENTRY ENTER.
(18) Press DATA ENTRY 2 key (2.5 V calibration), use DATA SCROLL controls to
adjust multimeter indication for 2.500 .010 V ac then press DATA ENTRY ENTER key.
(19) Press ESC (F6) key, press FIELD SELECT ⇑ or ⇓ key to highlight 8. VRMS
Meter, and press DATA ENTRY ENTER key.
16