TB 9-6625-2296-35
(20) Press DATA ENTRY ENTER and use DATA ENTRY keys to enter multimeter
indication to two decimal places, then press DATA ENTRY ENTER and Ret (F5) keys.
(21) Press MODE AF GEN (AF GEN Display) key; when prompted for backup data,
press SHIFT and Y keys (Backup Cal Data ? y/n).
(22) Disconnect test equipment, press TI POWER key off, remove frequency
proper slot. Press TI POWER key ON.
a. Performance Check
(1) Press key sequence listed in (a) through (f) below:
Meters Menu
(a) MODE MTRS .........................
Auxiliary Functions Menu
(b) AUX (F6) .................................
Recall Parameters Menu
(c) MEMORY RCL ......................
(d) FIELD SELECT ⇑ .................
Highlight 10. Factory Defaults
Factory Defaults recall
(e) DATA ENTRY ENTER .........
NOTE
Assure highlighted area indicates YES before proceeding.
Factory Defaults restored
(f) DATA ENTRY ENTER .........
(2) Connect equipment as shown in figure 3. Set audio analyzer source level to
minimum, multimeter to measure volts ac, and press calibrator WIDEBAND key on.
Figure 3. SINAD/distortion accuracy connection.
(3) Press key sequence listed in (a) through (m) below:
Meter Menu
(a) MODE MTRS ..........................
Dist Meter
(b) DATA ENTRY 6 ......................
Distortion Meter Menu
(c) SETUP .....................................