TB 9-6625-2296-35
Scope/Analyzer Menu
(b) SETUP .....................................
Scope On
(c) DATA ENTRY 1 ......................
Setup Scope
(d) DATA ENTRY 3 ......................
Input
(e) DATA ENTRY 1 ......................
AC
(f) DATA ENTRY 7 ......................
Trigger Source Internal
(g) DATA ENTRY 2 ......................
Sweep Rate
(h) DATA ENTRY 3 ......................
Highlight 1 ms
(i) DATA SCROLL ⇓ ...................
1 ms sweep
(j)
DATA ENTRY ENTER ..........
Return
(k) Ret (F5) ....................................
(3) Connect calibrator OUTPUT HI and LO to TI SCOPE IN.
(4) For TI scale settings listed in table 11, press Scale (F1) key, adjust DATA
SCROLL knob to select TI scale settings, press DATA ENTRY ENTER key, set calibrator
frequency to 1 kHz, and amplitude for 5 divisions of vertical deflection on TI. Calibrator
amplitude will be within limits specified in table 11.
NOTE
If necessary, press Vert (F2) key and use DATA SCROLL
controls to position waveform vertically on display.
Test instrument
Calibrator indications
Scale settings
Min
Max
4.1006
mV
2
mV
2.9694 mV
9.5445
mV
5
mV
8.1305 mV
19.089
mV
10
mV
16.261
mV
38.178
mV
20
mV
32.522
mV
95.445
mV
50
mV
81.305
mV
190.89
mV
100
mV
162.61
mV
381.78
mV
200
mV
325.22
mV
954.45
mV
500
mV
813.05
mV
1.9089
V
1
V
1.6261 V
3.8178
V
2
V
3.2522 V
9.5445
V
5
V
8.1305 V
19.089
V
10
V
16.261
V
38.178
V
20
V
32.522
V
95.445
V
50
V
81.305
V
(5) Connect oscilloscope calibrator SOURCE/MEASURE CHAN 1 to TI SCOPE
IN using 50 Ω feedthrough termination.
(6) Press Sweep (F3) key, adjust DATA SCROLL knob to select 20 s, and press
DATA ENTRY ENTER key.
31