TB 9-6625-2296-35
(6) If displayed signal peak is not centered on center vertical graticule line, perform
b (1) through (11) below.
(7) Set measuring receiver to measure frequency. Enter signal generator inital
position TI display signal peak on center vertical graticule line, set measuring receiver to
measure FREQ ERROR, and press MHz (INPUT FREQ) key.
second vertical graticule line. If measuring receiver does not indicate within limits
specified in table 13, perform b (1) through (11) below.
(9) Increase signal generator frequency until displayed signal peak is centered on
tenth vertical graticule line. If measuring receiver does not indicate within limits specified
in table 13, perform b (1) through (11) below.
Signal
Measuring receiver indications
Test instrument
generator No. 1
Min
Max
Settings
Graticule
initial setting
(kHz)
(kHz)
scan
indication
10 MHz
-3.5
-4.5
1 kHz
2d
10 MHz
3.5
4.5
1 kHz
10th
10 MHz
-7.0
-9.0
2 kHz
2d
10 MHz
7.0
9.0
2 kHz
10th
20 MHz
-17.5
-22.5
5 kHz
2d
20 MHz
17.5
22.5
5 kHz
10th
20 MHz
-35.0
-45.0
10 kHz
2d
20 MHz
35.0
45.0
10 kHz
10th
20 MHz
-70.0
-90.0
20 kHz
2d
20 MHz
70.0
90.0
20 kHz
10th
100 MHz
-175.0
-225.0
50 kHz
2d
100 MHz
175.0
225.0
50 kHz
10th
100 MHz
-350.0
-450.0
100 kHz
2d
100 MHz
350.0
450.0
100 kHz
10th
100 MHz
-700.0
900.0
200 kHz
2d
100 MHz
700.0
900.0
200 kHz
10th
2001 MHz
-1750.0
-2250.0
500 kHz
2d
200 MHz
1750.0
2250.0
500 kHz
10th
200 MHz
-2500.0
-3500.0
1 MHz
3d
200 MHz
2500.0
3500.0
1 MHz
9th
Set measuring receiver TRACK MODE on.
1
(10) Press FIELD SELECT keys to select Freq and Scan functions and press
DATA SCROLL arrow keys. Adjust DATA SCROLL knob to set Scan and Freq
remaining values. Repeat (3) through (5) and (7) through (9) above for remaining settings
(11) Press key sequence listed in (a) through (g) below:
Meters Menu
(a) MODE MTRS ..........................