TB 9-6625-2297-35
IDENTIFICATION AND DESCRIPTION
1. Test Instrument
Identification.
This bulletin provides instructions for the
calibration of Power Meter, Hewlett-Packard, Model 437B. The manufacturer's manual
was used as the prime data source in compiling these instructions. The equipment being
calibrated will be referred to as the TI (test instrument) throughout this bulletin.
a. Model Variations. None.
b. Time and Technique. The time required for this calibration is approximately 2
hours, using the dc and low frequency technique.
2. Forms,
Records, and
Reports
a. Forms, records, and
reports required for calibration personnel at all
levels are
prescribed by TB 750-25.
b. Adjustments to be reported are designated (R) at the end of the sentence in which
they appear. Report only those adjustments made and designated with (R).
3. Calibration
Description. TI
parameters and
performance specifications which
pertain to this calibration are listed in table 1.
Test instrument
parameters
Performance specifications
Zero set (digital set
0.5% FS (Most sensitive range. Decrease percentage factor of 10 for each higher
ability of zero)
range 1 count)
Instrument accuracy
Range: 3 W to 100 mW
single channel mode
Accuracy:1 0.5% or 0.2 dB (within same calibration range)
Frequency: 50 MHz
Accuracy: 0.5 MHz
Power reference
Range: 1 mW
Accuracy: 1.2%
Frequency: 50 MHz
includes sensor linearity. When operating in ranges 4 or 5 add the corresponding sensor power linearity
percentage.