TB 9-6625-2297-35
EQUIPMENT REQUIREMENTS
4. Equipment
Required. Table 2 identifies the
specific equipment to be
used in this
calibration procedure. This equipment is issued with Secondary Transfer Calibration
Standards Set AN/GSM-287, or AN/GSM-705. Alternate items may be used by the
calibrating activity. The items selected must be verified to perform satisfactorily prior to
use and must bear evidence of current calibration. The equipment must meet or exceed the
minimum use specifications listed in table 2. The accuracies listed in table 2 provide a four-
to-one ratio between the standard and TI. Where the four-to-one ratio cannot be met, the
actual accuracy of the equipment selected is shown in parenthesis.
5. Accessories Required. The
accessories required for this
calibration are common usage
accessories issued as indicated in paragraph 4 above and are not listed in this calibration
procedure. The following peculiar accessory is also required for this calibration: Power
sensor cable, Hewlett-Packard, Model 11730A.
Table 2. Minimum Specifications of
Equipment Required
Minimum use
Manufacturer and model
Common name
specifications
(part number)
FREQUENCY
Fluke, Model PM6681/656
COUNTER
Accuracy: 0.25%
(PM6681/656)
Ability to measure approximately 200Ω
Hewlett-Packard, Model
3458A (3458A)
Ability to measure V dc
POWER METER
Range: 0.988 to 1.012 mW
Hewlett-Packard, Model E12-
Accuracy: 0.7%
432A (MIS-30525)
w/thermistor mount, Hewlett-
Must have VCOMP and VRF outputs
Packard, Model H75-478A
(7915907)
RANGE
Range: 3 W to 100 mW
Hewlett-Packard, Model
CALIBRATOR
11683A (11683A)
Accuracy: 0.25%
6. Preliminary
Instructions
a. The
instructions outlined in
paragraphs 6 and
7 are preparatory to the
calibration
process. Personnel should become familiar with the entire bulletin before beginning the
calibration.
b. Items of equipment used in this procedure are referenced within the text by common
c. Unless otherwise specified, verify the result of each test and, whenever the test
requirement is not met, take corrective action before continuing with the calibration.
Adjustments required to calibrate the TI are included in this procedure. Additional
maintenance information is contained in the manufacturer's manuals.
d. Unless otherwise specified, all controls and control settings refer to the TI.
CHANGE 1 3