TB 9-6625-2304-24
EQUIPMENT REQUIREMENTS
4. Equipment Required. Table 2 identifies the specific equipment to be used in this
calibration procedure. This equipment is issued with Secondary Transfer Calibration
Standards Sets AN/GSM-286, AN/GSM-287 or AN/GSM-705. Alternate items may be used
by the calibrating activity. The items selected must be verified to perform satisfactorily
prior to use and must bear evidence of current calibration. The equipment must meet or
exceed the minimum use specifications listed in table 2. The accuracies listed in table 2
provide a four-to-one ratio between the standard and TI. Where the four-to-one ratio
cannot be met, the actual accuracy of the equipment selected is shown in parenthesis.
5. Accessories Required. The accessories required for this calibration are common
usage accessories, issued as indicated in paragraph 4 above and are not listed in this
calibration procedure. The following peculiar accessories are also required for this
calibration: 6 each, banana-to-pin adapters; Dc power supply, Elgar, Model
D C S 4 0 - 30EM10 (13589313).
Table 2. Minimum Specifications of Equipment Required
Manufacturer and model
Common name
Minimum use specifications
(part number)
Fluke, Model 5720A (5720A)
CALIBRATOR
Dc voltage output: 1.0 V
1
(p/o MIS-35947)
Accuracy:
Dc current output: 1.0 A
Accuracy:2
Ac voltage capability:
Output: 1.0 V
Agilent, Model 3458A (3458A)
MULTIMETER
Dc voltage:
Range: 6.0 mV to 18 V
Accuracy: (%)
6.0 mV1
Input:
10 to 60 mV2
90 mV to 9.0 V: 0.016
18 V: 0.0056
Ac voltage:
Range: 1.0 V and 10 V
Frequency: 20 kHz
Accuracy: 0.03%.
Range: 100 to 1.0 M
Accuracy: 0.0025%
Range: 20 kHz
Accuracy: 0.25%
1Combined
accuracy of calibrator (1.0 V dc) and multimeter (6.0 mV dc, nominal) is +0.025%.
2Combined
accuracy of calibrator (dc current) and multimeter (10 to 60 mV dc) is 0.125%.