SECTION III. CALIBRATION PROCESS
Ac Voltage and Frequency

Calibration Procedure For A31U12661-3, Self-Alignment Interface Connection Device Page Navigation
  1    2  3  4    5    6    7  
img
. .
TB 9-6625-2304-24
Figure 1. Dc and ac voltage.
Table 3. Dc Voltage
Multimeter indications
INPUT HI
(V)
to
test instrument
Min
Max
TP25
0.89550
0.90450
TP24
8.95500
9.04500
TP22
17.9100
18.0900
b. Adjustments. Ensure that dc power supply output is 36 (1.0) V dc; otherwise, no
adjustments can be made.