• Home
  • Download PDF
  • Order CD-ROM
  • Order in Print
Load Resistance (10 to 60 m )

Calibration Procedure For A31U12661-3, Self-Alignment Interface Connection Device Page Navigation
  2    3    4    5    6  7
img
. .
TB 9-6625-2304-24
(1) Connect TI's thermal converter output (P2/J2 of TC1) (after being disconnected
from internal connections) to multimeter and TI thermal converter input (P1/J1 of TC1)
to calibrator.
(2) Set calibrator for a 1.0 V dc output. Allow approximately 30 minutes for thermal
converter stabilization.
NOTE
Thermal converter in (2) above should be covered, so as to
shield from changing air currents.
(3) Record multimeter indication; then reverse polarity of calibrator output. Record
multimeter indication again. If the two indications are not equal, average them and record
value as DC Reference.
(4) Set calibrator frequency to 25 Hz.  Using output adjustment control, adjust
calibrator amplitude for a multimeter indication equal to value recorded as DC Reference
(0.5 V). Calibration Error indication will not exceed + 1.0%.
(5) Repeat technique of (4) above for 50 Hz, 20 kHz, 60 kHz, 175 kHz and 750 kHz.
Calibrator Error indication will not exceed + 1.0%.
b. Adjustments. No adjustments can be made.
13. Final Procedure
a. Deenergize and disconnect all equipment.
b. Annotate and affix DA label/form in accordance with TB 750-25.
9/(10 Blank)


Privacy Statement
Press Release
Contact

© Copyright Integrated Publishing, Inc.. All Rights Reserved.