TB 9-6625-2315-35
EQUIPMENT REQUIREMENTS
4. Equipment
Required.
Table 2 identifies the
specific equipment to be
used in this
calibration procedure. This equipment is issued with Secondary Transfer Calibration
Standards Set AN/GSM-286. Alternate items may be used by the calibrating activity. The
item selected must be verified to perform satisfactorily prior to use and must bear evidence
of current calibration. The equipment must meet or exceed the minimum use specifications
5. Accessories Required. The
accessories required for this
calibration are common
usage accessories, issued as indicated in paragraph 4 above, and are not listed in this
Table 2. Minimum Specifications of
Equipment Required
Manufacturer and model
Common name
Minimum use specifications
(part number)
CALIBRATOR
Dc voltage:
John Fluke, Model 5720A (p/o MIS-
Range: 100 mV to 1000 V
35947); w/power amplifier, John Fluke,
Accuracy: 0.001%
Model 5725A (5725A)
Range
Accuracy
0Ω
≤0.001 Ω
100 Ω to 1 MΩ
0.002 %
10 MΩ
0.010 %
100 MΩ
0.200 %
Dc current:
Range
Accuracy
0.5 A
0 mA
0.017 %
10 mA
0.013 %
100 mA
0.027 %
1A
0.037 %
2A
Ac voltage:
Range: 100 mV to 750 V
Frequency
Accuracy
0.022 %
10 Hz to 20 kHz
0.042 %
20 to 50 kHz
0.17 %
100 kHz
1.12 %
300 kHz
Ac Current:
.035 %
1A
1 kHz
.052 %
2A
1 kHz
FUNCTION/ARITRARY
Function: Sinewave
Agilent, Model 33250A (MIS-45853)
Amplitude: 1 Vrms
Accuracy: 0.0025 %