TB 9-6625-2315-35
a. Performance Check
(1) Connect calibrator OUTPUT HI and LO to TI Input HI and LO.
(2) Press keys as listed in (a) through (e) below:
(a) FUNCTION AC V.
(b) Shift then MENU <.
(c) RANGE/DIGITS ∨ until AC FILTER is displayed.
(d) MENU < or > until SLOW 3 HZ is displayed.
(e) RANGE/DIGITS Auto/Man.
(3) Press Shift key then RANGE/DIGITS Auto/Man key to select 6 digits
resolution.
(4) Press RANGE/DIGITS ∧ or ∨ key to select TI ranges and set calibrator output
for settings listed in table 9. If TI does not indicate within limits specified, perform b
below.
Test instrument
Calibrator output settings
Test instrument indications
ranges
Voltage
Min
Max
100
mV
100
mV
20
Hz
99.9
mV
100.1
mV
100
mV
100
mV
15
kHz
99.9
mV
100.1
mV
100
mV
100
mV
40
kHz
99.83
mV
100.17
mV
100
mV
100
mV
90
kHz
99.32
mV
100.68
mV
100
mV
100
mV
300
kHz
95.5
mV
104.5
V
1
V
1
V
20
Hz
0.9991
V
1.0009
V
1
V
1
V
15
kHz
0.9991
V
1.0009
V
1
V
1
V
40
kHz
0.9983
V
1.0017
V
1
V
1
V
90
kHz
0.9932
V
1.0068
V
1
V
1
V
300
kHz
0.955
V
1.045
V
10
V
10
V
10
Hz
9.991
V
10.009
V
10
V
10
V
15
kHz
9.991
V
10.009
V
10
V
10
V
40
kHz
9.983
V
10.017
V
10
V
10
V
90
kHz
9.932
V
10.068
V
10
V
10
V
300
kHz
9.55
V
10.45
V
100
V
100
V
20
Hz
99.91
V
100.09
V
100
V
100
V
15
kHz
99.91
V
100.09
V
100
V
100
V
40
kHz
99.83
V
100.17
V
100
V
100
V
90
kHz
99.32
V
100.68
V
100
V
100
V
200
kHz
95.5
V
104.5
V
750
V
750
V
40
Hz
749.325
V
750.675
V
750
V
750
V
90
kHz
744.9
V
755.1
V
CHANGE 2