TB 9-6625-2317-40
(7) Connect equipment as shown in figure 6, CONNECTION B.
(8) Press TI pushbuttons (on) as indicated in (a) through (d) below:
(a) CARR FREQ then 1, 0, 0, MHz.
(b) LEVEL DISPLAY to FM.
(c) LEVEL DETECTOR MODE to PK-PK/2.
(d) LOW PASS FILTER to 300 kHz.
(9) TI LEVEL/DATA ENTRY display will indicate within 3% +1 count of the
recorded value in (6) above.
(10) Connect equipment as shown in figure 6, CONNECTION A.
(11) Set audio analyzer SOURCE LEVEL to 764 mV.
(12) On measuring receiver, record frequency modulation displayed indication.
(13) Connect equipment as shown in figure 6, CONNECTION B.
(14) TI LEVEL/DATA ENTRY display will indicate within 3% +1 count of the
recorded value in (12) above.
12