TB 9-6625-2321-35
IDENTIFICATION AND DESCRIPTION
1. Test Instrument
Identification.
This bulletin provides instructions for the
calibration of Digital Multimeter, National Instruments, Model DAQCard-4050. The
manufacturer's manual was used as the prime data source in compiling these instructions.
The equipment being calibrated will be referred to as the TI (test instrument) throughout
this bulletin. The AN/PSM-95 or AN/PSM-95A and current shunt, National Instruments
CSM-200ma (part of TI) will be required for this procedure.
a. Model Variations. None.
b. Time and Technique. The time
required for this
calibration is
approximately 1
hour, using the dc and low frequency technique.
2. Forms,
Records, and
Reports
a.
Forms, records, and
reports required for calibration personnel at all
levels are
prescribed by TB 750-25.
b. Adjustments to be reported are designated (R) at the end of the sentence in which
they appear. When adjustments are in tables the (R) follows the designated adjustment.
Report only those adjustments made and designated with (R).
3. Calibration
Description. TI
parameters and
performance specifications which
pertain to this calibration are in table 1.
Test instrument
parameters
Performance specifications
Dc voltage
Range: .2 to 200 V
Accuracy: 1% of range
Ac voltage
Range: .2 to 200 V
Accuracy: 3% of range
Range: 0 to 20 MΩ
Accuracy: 1% of range
Dc current1
Range: 20 to 200 mA dc
Accuracy: 1% of range
Ac current1
Range: 20 to 200 mA ac
Accuracy: 3% of range
1Current
measurements require current shunt (CSM-200ma).