TB 9-6625-2321-35
Calibrator
Range
Test instrument indications
output settings
pushbutton
@ 1000 Hz
(V)
settings
(V)
(V)
Min
Max
200 m
190 m
184.00 m
196.00 m
2
1.90
1.8400
1.9600
25
19.0
18.400
19.600
250
190
184.00
196.00
b. Adjustments. Perform the procedure as outlined in 8 b above.
a. Performance Check
Set calibrator output and TI controls to values listed in table 5. If TI indications are
not within the limits in table 5, perform b below.
Range
Calibrator
pushbutton
output
Test instrument indications
settings
settings
Ω
Min
Max
Ω
Ω
200
190
188.00
192.00
2k
1.90 k
1.8800
k
1.9200
k
20 k
19.0
k
18.800
k
19.200 k
200 k
190
k
188.00 k
192.00 k
2M
1.90 M
1.8800
M
1.9200
M
20 M
19.0 M
18.800
M
19.200
M
b. Adjustments. Perform the procedure as outlined in 8 b. above.
a. Performance Check
(1) Disconnect TI red (+) and black (-) test probes from calibrator OUTPUT HI and
LO terminals.
(2) Connect current shunt (CSM-200 ma) to TI red (+) and black (-)test probes.
(3) Connect calibrator OUTPUT HI and LO terminals to current shunt (CSM-200
ma) red (+) and black (-) connection.
(4) Click on Edit menu at the top of the DMM display.