TB 9-6625-2322-35
IDENTIFICATION AND DESCRIPTION
1. Test Instrument
Identification.
This bulletin provides instructions for the
used as the prime data source in compiling these instructions. The equipment being
calibrated will be referred to as the TI (test instrument) throughout this bulletin.
a. Model Variations. None.
b. Time and Technique. The time required for this calibration is approximately 5
hours, using the dc and low frequency and microwave techniques.
2. Forms,
Records, and
Reports
a.
Forms, records, and
reports required for calibration personnel at all
levels are
prescribed by TB 750-25.
b. Adjustments to be reported are designated (R) at the end of the sentence in which
they appear. When adjustments are in tables, the (R) follows the designated adjustment.
Report only those adjustments made and designated with (R).
3. Calibration
Description. TI
parameters and
performance specifications which
pertain to this calibration are listed in table 1.
Test instrument parameters
Perform specifications
Range: 10 MHz to 18 GHz
Accuracy: < 5 parts in 10-10
Time Base Stability: < 5 parts in 10-10 per day
Harmonics
10 MHz to ≤ 50 MHz
<-30 dBc
<-40 dBc
> 50 MHz to ≤ 2 GHz
<-60 dBc
> 2 GHz to ≤ 18 GHz
<-40 dBc
10 MHz to ≤ 2 GHz
<-60 dBc
> 2 GHz to ≤ 18 GHz
RF output
Accuracy: 1 dB
Level: 11 dBm
Flatness:
Accuracy:
10.0 MHz to 50.0 MHz
2 dB for a 11 dB output level
50 MHz to 18 GHz
0.8 dB for a 11 dB output level
Attenuator:
0 to -122 dB1
Accuracy: 1 dB
See footnotes at end of table.