TB 9-6625-2322-35
Adjustments required to calibrate the TI are included in this procedure. Additional
maintenance information is contained in the manufacturer's manual for this TI.
d. Unless otherwise specified, all controls and control settings refer to the TI.
WARNING
HIGH VOLTAGE is used or exposed during the performance of
DEATH ON CONTACT may result if
personnel fail to observe safety precautions.
REDUCE
OUTPUT(s) to minimum after each step within the
performance check where applicable.
a. Remove TI from protective cover only as necessary to make adjustments.
NOTE
For the remainder of this procedure the SG-1219/U connected
to the synthesized signal generator workstation will be called
CAUTION
Before connecting TI to power source, make sure TI is set to the
power source line voltage as shown on rear of TI.
b. Connect TI to 115 V ac power source.
c. Set LINE switch to ON and allow a 1 hour warm-up and stabilize.
e. Remove screw covering the 10 MHz REF ADJ
(fig. 2)
f. Adjust 10 MHz REF ADJ
(fig. 2) for a minimum frequency difference
meter indication.
g. After 24
hours, verify that
frequency difference meter drift
indication is
less than 5
parts in 10-10 per day.
i. Replace TI top cover.
j. Disconnect frequency
difference
meter
from
the
TI
and
the