TB 9-6625-2322-35
(3) Disconnect the measuring receiver power sensor module from
the
CALIBRATION RF POWER OUTPUT and
connect it to the TI
RF OUTPUT.
(4) Press TI keys as listed in (a) through (e) below:
(a) SYSTEM, Reset.
(b) OUTPUT off.
(c) Frequency Control, F0, Edit F0, 2, and GHz.
(d) OUTPUT on.
(e) Level Control, L0, Edit L0, 0, and dB.
(5) Using measuring receiver and RF power techniques in Log Mode verify that the
measuring receiver indicates within tolerances listed in table 5.
(6) Using technique of (4) (e) set the TI to the remaining levels listed in table 5 and
repeat (5) above.
Table 5. 2 GHZ
Output Level Test 1 dB
Steps
Test instrument
output level
(dB)
Min
Max
0
-1
1
-1
-2
0
1
0
2
2
1
3
3
2
4
4
3
5
5
4
6
6
5
7
7
6
8
8
7
9
9
8
10
10
9
11
11
10
12
(7) Press Level Control, L0, Edit L0, 0, and dB keys.
(8) Using standard tuned level measurement techniques, verify the measuring
receiver indicates within minimum and maximum limits for TI output level as listed in
(9) Use the TI arrow key to decrement the output level 10 dB as indicated in table 6
verifying that the indication is within limits listed in table 6.
(10) Repeat (9) above for remaining TI output level settings listed in table 6.
10