TB 9-6625-2322-35
(16) Use the TI arrow key to decrement the output level 10 dB as indicated in table 8
and repeat (15) above.
(17) Repeat (16) above for remaining TI output level settings listed in table 8.
Table 8. 5 GHz
Output Level Test 10 dB
Steps
Test instrument
output level
(dB)
Min
Max
0
-1
1
-10
-11
-8
-20
-21
-18
-30
-31
-28
-40
-41
-38
-50
-51
-48
-60
-61
-58
-70
-71
-66.5
-80
-81
-76.5
-90
-91
-86.5
(18) Reduce TI output to minimum.
(19) Disconnect TI from measuring receiver.
b. Adjustments. No adjustments can be made.
a. Performance Check