TB 9-6625-2322-35
(14)
Reduce all outputs to minimum.
(15)
Disconnect equipment setup.
b. Adjustments. No further adjustments can be made.
a. Performance Check
(1) Connect equipment as shown in figure 6 below.
(2) Set measuring receiver to measure frequency modulation, with +PEAK detector,
<20 Hz high pass and >200 kHz lo pass filters and at a frequency of 5 GHz.
(3) Press TI keys as listed in (a) through (j) below.
(a) SYSTEM, Rest.
(b) OUTPUT off.
(c) Level Control, L0, Edit L0, 0, and dB.
(d) Frequency Control, F0, Edit F0, 5, and GHz.