TB 9-6625-2322-35
b. Adjustments. No further adjustments can be made.
a. Performance Check
(1) Connect equipment as shown in figure 7 below.
(2) Press TI keys as listed in (a) through (k) below.
(a) SYSTEM, Rest.
(b) OUTPUT off.
(c) Level Control, L0, Edit L0, 1, 1, and dB.
(d) Frequency Control, F0, Edit F0, 5, and GHz.
(e) OUTPUT on.
(f) Modulation, Pulse, and On/Off to on.
(g) More, Config, Clock 40/10 to 40.
(h) Top Menu, More, Mode, press arrow key to highlight Single, Select, and
Previous Menu.
(i) Trigger, press arrow key to highlight Free Run, Select, and Previous Menu.
(j)
Previous Menu, Edit Period, 1, 0, uS, Edit Period.
(k) Edit Width, 5, uS, and Edit Width.
(a) Vertical 1 Input 50Ω to on (lit).
(b) Vertical 1 scale to 20 mV.