TB 9-6625-2323-35
5. Accessories Required. The
accessories required for this
calibration are common
usage accessories, issued as indicated in paragraph 4 above, and are not listed in this
Detector, Hewlett-Packard, Model HP-423A.
Table 2. Minimum Specifications of
Equipment Required
Manufacturer and model
Common name
Minimum use specifications
(part number)
AUDIO ANALYZER
Range: 1.0 to 100.0 kHz
Boonton, Model 1121
Accuracy: <2.0%
(1121)
FREQUENCY
Range: 10 MHz
Tracor, Model 527E (MIS-
DIFFERENCE METER
Resolution: 1 part in 10
10318)
-8
MEASURING RECEIVER
Range: 2.0 to 18 GHz
Hewlett-Packard, Model
Range: +11 to -74.5 dBm
8902A w/sensors, Hewlett-
Accuracy: +.5 dB
Packard, Model 11792A
Range: AM 0 to 80%
(11792A), and microwave
Accuracy: 2% at 1 kHz
converter, model 11793A
(11793A)
Range: FM .05 to 100 kHz
Accuracy: 2% at 1 kHz
Deviation: 12 kHz
MICROWAVE
Range: 10 MHz to 18 GHz
Hewlett-Packard Model
FREQUENCY COUNTER
Accuracy:
5352B-001 (5352B-001)
1
Range: 5.0 V at 100 ns
(OS-303/G)
Accuracy: 3.0%
Anritsu, Model 68369NV
receiver
(68369NV)
(AN/USM-677)
SPECTRUM ANALYZER
Range: 2 to 18 GHz at +10 to 60 dBm
+20% of the span/div
Range: Span 500 Hz to 20 MHz
5%
Accuracy:
TIME/FREQUENCY
Frequency: 1 MHz
Datum, Model ET6000-75
WORKSTATION
Accuracy: 5 parts in 10-10 per day
(13589305)
1Time
base tied to the TI.
6. Preliminary
Instructions
a. The
instructions outlined in
paragraphs 6 and
7 are preparatory to the
calibration
process. Personnel should become familiar with the entire bulletin before beginning the
b. Items of equipment used in this procedure are referenced within the text by common