TB 9-6625-2323-35
NOTE
TI must have power applied and in STANDBY or OPERATE
for at least 120 continuous hours before the frequency drift
specification can be verified.
e. Verify that the
frequency difference meter drift is
less than 5
parts in 10
-8 per day.
f.
Disconnect frequency difference meter from the TI
and the time/frequency workstation.
a. Performance Check
(1) Connect equipment as indicated in figure 2.
(2) Press TI keys as listed in (a) through (e) below.
(a) System, Reset.
(b) Output On Off to off.
(c) Level Control, L0, Edit L0, 0, and MHz ms dB.
(d) Frequency Control, F0, Edit F0, 1, and GHz Sec dBm.
(e) Output On Off to on.
(3) Set microwave frequency counter to measure frequency on INPUT 1.
(4) Verify that the microwave frequency counter indicates within limits listed in