TB 9-6625-2323-35
Table 9. GHz
Output Level Test 1 dB
Steps
Test instrument
output level
(dB)
Min
Max
0
-1
1
-1
-2
0
1
0
2
2
1
3
3
2
4
(21) Press Level Control, L1, Edit L1, 0, and MHz ms dB.
(22) Using standard tuned level measurement techniques, verify the measuring
receiver indicates within minimum and maximum limits for TI output level as listed in
(23) Use the TI arrow key to decrement the output level as indicated in table 10 and
repeat (22) above.
(24) Repeat (23) above for remaining TI output level settings listed in table 10.
Table 10. 18 GHz
Output Level Test 10 dB
Steps
Test instrument
output level
(dB)
Min
Max
0
-1
1
-10
-11
-9
-20
-21
-19
-30
-31
-29
-40
-41
-39
-50
-51
-49
-60
-61
-59
-70
-71
-69
-80
-81
-79
-90
-91
-89
(25) Reduce TI output to minimum.
(26) Disconnect TI from measuring receiver.
b. Adjustments. No adjustments can be made.
a. Performance Check