TB 9-6625-2326-35
to-one ratio between the standard and TI. Where the four-to-one ration cannot be met, the
actual accuracy of the equipment selected is shown in parenthesis.
5. Accessories Required. The
accessories listed in table 3
are issued as
indicated in
paragraph 4 above and
are used in this
calibration procedure. When necessary, these items
may be substituted by equivalent items, unless specifically prohibited.
Table 2. Minimum Specifications of
Equipment Required
Manufacturer and model
Common name
Minimum use specifications
(part number)
CALIBRATOR
Range: 0.2 V dc to 200 V dc
Fluke, Model 5720A (5700A/EP)
Accuracy: 0.25%
(p/o MIS-35947); w amplifier,
Fluke 5725A/AR (5725A/AR)
Range: 0.2 V ac to 200 V ac
Accuracy: 0.75%
Range: 0 to 20 MΩ
Accuracy: 0.25%
Common name
Description (part number)
CONTROLLER 1
Polywell (MIS-45854)
Miltope, AN/PSM-95
SPORT or MSD
1
Miltope, AN/PSM-95A
(To be provided by customer for
(Turned in with TI if required)
calibration of DM224 if Polywell
Controller w/preinstalled
calibration software not available.)
Calibration software
SPORT/MSD DMM
(provided to teams by PM-TMDE)
Maintenance Disk
1
(Provides calibration software
routine for use on SPORT/MSD.)
Test Lead Connector 2
Type 311 cable
1Verify
you have latest calibration software updates and other necessary information
from USATA homepage at TB 9-6625-2326-35 folder location.
2To be supplied with TI.
(UTILIZING POLYWELL CONTROLLER)
NOTE
Use of this section of the TB requires use of the Polywell
Controller (MIS-45854) with the DM224 Calibration Software
pre-installed. If you do not have access to this controller, you
will need to perform Section IV. Be advised that this pre-
installed software is unique and not to be confused with the
SPORT/MSD DMM Maintenance Disk that is used in Section IV.
CHANGE 1