TB 9-6625-2326-35
b. Adjustments
The test and calibration utility will give pass/fail indications. If there is an out-of-
tolerance finding on any function/range of the Performance Check, you will be given an
opportunity to repeat the test. If the out-of-tolerance finding still exists, you will be
prompted to run the Alignment Check. Follow on-screen instructions. When the Alignment
Check is successfully completed, you must rerun the Performance Check for a successful
completion of same.
a. Deenergize and disconnect all equipment.
b. Annotate and affix DA label/form in accordance with TB 750-25.
UTILIZING AN/PSM-95 (SPORT) OR AN/PSM-95A (MSD)
NOTE
Use this section of TB if calibrating the DM224 utilizing the
"SPORT/MSD DMM Maintenance Disk" running from the
SPORT (AN/PSM-95) or MSD (AN/PSM-95A) CD drive.
10. Preliminary
Instructions
a. The
instructions outlined in
paragraphs 10 and
11 are preparatory to the
calibration
process. Personnel should become familiar with the entire bulletin before beginning the
calibration.
b. Items of equipment used in this procedure are referenced within the text by common
c. Unless otherwise specified, verify the result of each test and, whenever the test
requirement is not met, take corrective action before continuing with the calibration.
Adjustments required to calibrate the TI are included in this procedure. Additional
maintenance information is contained in the manufacturer's manual for this TI.
d. Unless otherwise specified, all controls and control settings refer to the TI.
e. Unless otherwise specified, on screen controls will be accessed by using the mouse or
the trackball controls to move the arrow over the icon, virtual instrument controls, etc., and
clicking the left button.