TB 9-6625-2327-35
(3) Set frequency counter to measure frequency on channel A.
(4) If the frequency counter does not indicate within limits listed in table 3, perform
(5) Set TI frequency to the next frequency listed in table 3, using the TI key pad and
control knob, and repeat (4) above.
(6) Repeat (2) (c), (4) and (5) above for the remaining frequencies listed in table 3.
Microwave frequency counter indications
Test instrument
center frequency
Min
Max
10.000000 MHz
9.999990
MHz
10.000010
MHz
20.000000 MHz
19.999980
MHz
20.000020
MHz
40.000000 MHz
39.999960
MHz
40.000040
MHz
80.000000 MHz
79.999920
MHz
80.000080
MHz
1.0000000 MHz
0.9999990 MHz
1.0000010 MHz
100.00000 kHz
99.99990
100.00010
kHz
kHz
10.000000 kHz
9.999990
kHz
10.000010
kHz
1.0000000 kHz
0.9999990 kHz
1.0000010 kHz
100.00000 Hz
0.00999999 s
0.01000001 s
1
10.000000 Hz
0.09999990 s
0.10000010 s
1.000000 Hz
0.99999900 s
1.00000100 s
100.000
mHz
9.9999900 s
10.0000100 s
1Set
TI to square wave, and frequency counter to measure period.
(7) Press Output to off.
(8) Disconnect microwave frequency counter from TI.
b. Adjustments.
Perform entire alignment procedure listed in
paragraph 13.
a. Performance Check
(1) Connect multimeter to the TI RF OUTPUT.
(3) Press TI keys as listed in (a) through (e) below:
(a)
Sine.
(b)
Utility, Output Setup, Load High Z to highlight High Z and Done.
(c)
Freq, 1, kHz.
(d)
Ampl, 10, and mV rms.
(e)
Output to on.
(4) If the multimeter does not indicate within minimum and maximum limits listed
(5)
Repeat technique of (3) and (4) above for remaining TI settings listed in table 4.