TB 9-6625-2327-35
Power meter indication
1kHz_10dB_reference
difference
Test instrument
Min
Max
200
kHz
-0.086
+0.087
500
kHz
-0.086
+0.087
1.500
MHz
-0.086
+0.087
5.000
MHz
-0.086
+0.087
10.000
MHz
-0.086
+0.087
25.000
MHz
-0.172
+0.175
40.000
MHz
-0.172
+0.175
50.000
MHz
-0.172
+0.175
60.000
MHz
-0.424
+0.446
65.000
MHz
-0.424
+0.446
70.000
MHz
-0.424
+0.446
75.000
MHz
-0.424
+0.446
80.000
MHz
-0.424
+0.446
NOTE
Use a 10 dB attenuator between TI and power meter.
Measured value will take attenuator into consideration during
calculations.
(9) Press TI Freq, 100, and kHz keys, and adjust the TI output level until the
power meter indication matches the calculated 100kHz_20dB_reference in dB minus the
attenuator test report value in dB at 100 kHz.
(10) Adjust the TI frequency to each of the frequencies listed in table 8. If the
difference between the power meter indication and the calculated 1kHz_20dB_reference
minus the attenuator test report value in dB at the selected frequencies is not within the
Power meter indication
1kHz_20dB_reference
difference
Test instrument
Min
Max
200
kHz
-0.086
+0.087
500
kHz
-0.086
+0.087
1.5 00 MHz
-0.086
+0.087
5.000
MHz
-0.086
+0.087
10.000
MHz
-0.086
+0.087
25.000
MHz
-0.172
+0.175
40.000
MHz
-0.172
+0.175
50.000
MHz
-0.172
+0.175
60.000
MHz
-0.424
+0.446
65.000
MHz
-0.424
+0.446
70.000
MHz
-0.424
+0.446
75.000
MHz
-0.424
+0.446
80.000
MHz
-0.424
+0.446
(11) Press Output to off and disconnect TI from power meter.
8