TB 9-6625-2327-35
Step #
Typical test requirements
8
-30 dB range with distortion filter
9
-20 dB range with distortion filter
10
-10 dB range with distortion filter
11
0 dB range with distortion filter
12
+10 dB range with distortion filter
13
-30 dB range without distortion filter
14
-20 dB range without distortion filter
15
-10 dB range without distortion filter
16
0 dB range without distortion filter
17
+10 dB range without distortion filter
(11) Disconnect the four-wire ohms connection and reconnect the TI Output to the
multimeter INPUT HI and LO. Set the multimeter to measure DCV.
(12) Press the BEGIN key, using the numerical keypad, adjust the displayed Meas'd
Voltage to
match the
multimeter indication, then
press the
ENTER VALUE key. Repeat
this technique for each setup listed in table 15. (Entered values are rounded to the nearest
100 V.)
Step #
Nominal DC level
Typical test requirements
18
+0.015 V
Output of 30 dB range
19
-0.015 V
Output of 30 dB range
20
+0.05
V
Output of 20 dB range
21
-0.05
V
Output of 20 dB range
22
+0.15
V
Output of 10 dB range
23
-0.15
V
Output of 10 dB range
24
+0.50
V
Output of 0 dB range
25
-0.50
V
Output of 0 dB range
26
+0.15
V
Output of 10 dB range (Amplifier in)
27
-0.15
V
Output of 10 dB range (Amplifier in)
28
+0.50
V
Output of 0 dB range (Amplifier in)
29
-0.50
V
Output of 0 dB range (Amplifier in)
30
+1.5
V
Output of +10 dB range (Amplifier in)
31
-1.5
V
Output of +10 dB range (Amplifier in)
32
+5
V
Output of +20 dB range (Amplifier in)
33
-5
V
Output of +20 dB range (Amplifier in)
(13) Set the multimeter to measure V rms.
(14) Press the BEGIN key, using the numerical keypad, adjust the displayed Meas'd
Vrms to
match the
multimeter indication, then
press the
ENTER VALUE key. Repeat
this technique for each setup listed in table 16.
13