TB 9-6625-2327-35
(6) Set the multimeter to display 5 digits and dc volts measurement.
(7) With Setup # 6 highlighted press the BEGIN key then using the numerical
keypad, adjust the displayed Meas'd Voltage to match the multimeter dc volt indication,
then press the ENTER VALUE key.
(8) Disconnect cable from the rear panel Modulation In and with Setup # 7
highlighted press the BEGIN key.
(9) Set the multimeter to measure offset-compensated, four-wire ohms, with 100
NPLC integration. Connect equipment as shown in figure 2.
Figure 2. Output impedance adjustment hook-up.
(10) Using the numerical keypad, adjust the displayed Meas'd Imped at each setup
to match the measured impedance then press the ENTER VALUE key. Repeat this
12